Specifications

All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise.

Voltage and current ranges
Voltage ranges
1000 Vpeak, 500 Vpeak, 200 Vpeak, 100 Vpeak, 50 Vpeak, 20 Vpeak, 10 Vpeak
Current ranges (20 A shunt)
100 Apeak, 50 Apeak, 20 Apeak, 10 Apeak, 5 Apeak, 2 Apeak, 1 Apeak, 0.5 Apeak, 0.2 Apeak, 0.1 Apeak
Current ranges (1 A shunt)
2.0 Apeak, 1.0 Apeak, 0.4 Apeak, 0.2 Apeak, 0.1 Apeak, 0.04 Apeak, 0.02 Apeak, 0.01 Apeak, 0.004 Apeak, 0.002 Apeak
Measurement accuracy – voltage
Voltage accuracy, VRMS (45 Hz to 850 Hz)
± 0.04% of Reading ± 0.04% of Range ± 0.005 V
Voltage accuracy, VRMS (10 Hz to 45 Hz, 850 Hz to 1 MHz, typical)
± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± 0.05 V
Voltage accuracy, DC (typical)
± 0.1% of Reading ± 0.1% of Range ± 0.05 V
Effect of common mode (typical)
100 V, 100 kHz
Measurement accuracy – current
Current accuracy, ARMS (45 Hz to 850 Hz) 1
± 0.04% of Reading ± 0.04% of Range ± (1.8 µV/Zext)
Current accuracy, ARMS (10 Hz to 45 Hz, 850 Hz to 1 MHz, typical)
± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± (50 µV/Zext)
Current accuracy, DC (typical)
± 0.1% of Reading ± 0.1% of Range ± (100 µV/Zext)
Current – peak inrush accuracy (100 Apeak range)
2% of Range ± 20 mA
Effect of common mode (typical)
100 V, 100 kHz, 20 A shunt

100 V, 100 kHz, 1 A shunt

100 V, 100 kHz, external shunt

1 Offset is valid in low bandwidth, with internal shunts, and after a manual zero has been performed. Offset is 10 µV/Zext in high bandwidth, and with external shunt.

Measurement accuracy – frequency
Frequency (10 Hz to 20 kHz)
0.1% of Reading, with the peak of the signal extending 10% above and 10% below the DC level
Frequency (20 kHz to 1 MHz)
0.1% of Reading, with the peak of the signal extending 25% above and 25% below the DC level
Measurement accuracy – power
Watts accuracy
± 0.075% of Reading ± 0.075% of Range (PF=1, 45 – 850 Hz)
VA accuracy
(Vrmsacc x Arms) + (Armsacc x Vrms)
VAR accuracy (typical)
Thumbnail
PF Accuracy
Cos θ -cos [ θ± ( Vh1 ph.err± Ah1 ph.err)] ± 0.002
Measurement accuracy – harmonic magnitude and phase (typical)
Voltage harmonics magnitude (10 Hz to 480 kHz)
± 0.02% of Reading ± 0.1% of Range ± (0.04*F)% of Reading ± 0.05 V
Voltage harmonics phase
± 0.04 ± [ 0.01 * (Vrange / Vreading )] ± (0.1 / Vrange) ± (0.005 *F )
Current harmonics magnitude (10 Hz to 480 kHz)
± 0.2% of Reading ± 0.1% of Range ± (0.04*F)% of Reading ± (50 µV / Zext)
Current harmonics phase
± 0.04 ± [0.01 * (Arange / Areading)] ± (0.001 / Arange * Zext) ± (0.005 *F)
Physical characteristics
Dimensions mm in
Height 102 4.0
Width 223 8.7
Depth 285 11.2
Weight Kg lb
Net (without lead set) 3.2 7.0
Temperature C F
Operating 0 °C to +40 °C +32 °F to +102 °F
Nonoperating -20 °C to +60 °C -4 °F to +140 °F

Notes:

All stated accuracies are based upon a minimum of a 30-minute warm-up period.

Zext is the external shunt impedance used and must be less than or equal to 10 Ohms.

If no frequency is measured, then the signal is considered DC for the purpose of accuracy.

F is the frequency measured in kHz. In the case of harmonics, F is the harmonic frequency.

Specifications are valid from 1 to 100% of range in low bandwidth and after a manual zero has been performed. Values below 1% are typical.

In high bandwidth, specifications are valid when the signal is greater than 10% of the range.

Harmonic specifications are always valid when the harmonic is greater than 2% of the range.

Measurement conditions during calibration: Instrument default settings unless otherwise stated, sine waves applied to V and I inputs, 30 min warm up, ambient temperature 23 °C ±5 °C.