Specifications
All specifications are warranted performance characteristics with guaranteed tolerance limits, unless noted otherwise. All specifications are valid between 18 °C and 28 °C ambient temperature and apply to all models unless noted otherwise.
Nominal refers to the features or characteristics of the design.
Typical indicates that 80% of units will meet the published typical performance with 80% confidence, unless otherwise noted. Typical performance is not warranted.
Typical-Mean indicates the mean performance measured on a sample of units immediately after performing an alignment. Typical-Mean performance is not warranted.
General
- Connectivity
- USB 2.0
- Software
- VectorVu-PC™ (requires Windows® 7/8/10, 64-bit)
Corrected system performance with calibration kit
-
- Spinner Type-N mechanical calibration kit BN533861
- User Correction ON.
Two 60 cm Tektronix cables 012-1765-00 or 012-1768-00
Frequency Directivity (dB)
Source match (dB)
Load match: Insertable devices (dB)
Load match with M-M or F-F thru (dB)
Reflection tracking (dB)
Transmission tracking: Insertable devices (dB)
300 kHz to < 1 MHz -38 -34 -37 -35 0.08 0.05 1 MHz to < 10 MHz -37 -34 -37 -35 0.08 0.02 10 MHz to < 100 MHz -37 -34 -37 -35 0.08 0.01 100 MHz to < 1 GHz -36 -34 -37 -35 0.08 0.01 1 GHz to < 3 GHz -36 -34 -37 -35 0.08 0.02 3 GHz to 6 GHz -36 -34 -36 -35 0.09 0.02 18 °C to 28 °C, within 1 °C of calibration temperature and at the same ambient humidity conditions in which calibration was performed. - Spinner 3.5 mm mechanical calibration kit BN533854
- User Correction ON.
Two 60 cm Tektronix cables 012-1769-00 or 012-1772-00
Frequency Directivity (dB)
Source match (dB)
Load match: Insertable devices (dB)
Load match with M-M or F-F thru (dB)
Reflection tracking (dB)
Transmission tracking: Insertable devices (dB)
300 kHz to < 1 MHz -36 -33 -37 -35.5 0.1 0.06 1 MHz to < 10 MHz -35 -33 -37 -35.5 0.1 0.02 10 MHz to < 100 MHz -35 -33 -37 -35.5 0.1 0.01 100 MHz to < 1 GHz -35 -33 -35 -35.5 0.1 0.01 1 GHz to < 4 GHz -35 -33 -35 -35.5 0.1 0.02 4 GHz to 6 GHz -28 -28 -29 -29 0.22 0.03 18 °C to 28 °C, within 1 °C of calibration temperature and at the same ambient humidity conditions in which calibration was performed. - Spinner Type-N calibration kit BN533844
- User Correction ON.
Two 60 cm Tektronix cables 012-1765-00
Frequency Directivity (dB)
Source match (dB)
Load match: Insertable devices (dB)
Load match with M-M or F-F thru (dB)
Reflection tracking (dB)
Transmission tracking: (dB)
300 kHz to < 1 MHz -32 -31 -32 -31 0.15 0.07 1 MHz to < 10 MHz -32 -31 -32 -31 0.15 0.03 10 MHz to < 100 MHz -31 -31 -32 -31 0.15 0.01 100 MHz to < 1 GHz -31 -31 -32 -31 0.15 0.01 1 GHz to < 4 GHz -31 -31 -32 -31 0.15 0.02 4 GHz to 6 GHz -25 -25 -26 -26 0.3 0.04 18 °C to 28 °C, within 1 °C of calibration temperature and at the same ambient humidity conditions in which calibration was performed.
Uncorrected system performance
- User correction OFF. Factory correction ON.
Frequency Directivity (dB)
Source match (dB)
Load match (dB)
Reflection tracking (dB)
Transmission tracking (dB)
300 kHz to < 500 kHz -25 -25 -4.5 ±1 ±1 500 kHz to < 2 MHz -25 -25 -4.5 ±1 ±1 2 MHz to < 10 MHz -25 -25 -11 ±1 ±1 10 MHz to < 200 MHz -25 -25 -11 ±1 ±1 200 MHz to <1.5 GHz -25 -25 -10 ±1 ±1 1.5 GHz to < 4.5 GHz -25 -25 -8 ±1 ±1 4.5 GHz to < 5 GHz -25 -25 -7 ±1 ±1 5 GHz to 6 GHz -25 -25 -7 ±1 ±1
Frequency
- Range
-
- TTR503A
- 100 kHz to 3.0 GHz
- TTR506A
- 100 kHz to 6.0 GHz
- Resolution
- 1 Hz
- Accuracy
- ± 7.0 ppm over 1 year calibration interval, 18 °C to 28 °C
- Internal reference
-
- Frequency
- 10 MHz
- Initial accuracy
- ±10 Hz
Aging
- External reference input
- 10 MHz ± 50 Hz
Test port output
- Dynamic range
-
Description 300 kHz to< 1 MHz 1 MHz to< 2 MHz 2 MHz to< 200 MHz 200 MHz to< 3 GHz 3 GHz to < 4.5 GHz
(TTR506A only)
4.5 GHz to 6 GHz (TTR506A only)
Dynamic range Typical 112 dB 117 dB 124 dB 125 dB 123 dB 122 dB Specification – – – 124 dB 123 dB 122 dB Noise floor -110 dBm/Hz
(< -115 dBm/Hz typical)< -120 dBm/Hz
(< -125 dBm/Hz typical)< -120 dBm/Hz
(< -125 dBm/Hz typical)< -125 dBm/Hz
(< -130 dBm/Hz typical)< -125 dBm/Hz
(< -130 dBm/Hz typical)< -125 dBm/Hz
(< -130 dBm/Hz typical)
- Crosstalk with load
-
Description 300 kHz to < 1 MHz
1 MHz to < 10 MHz
10 MHz to < 200 MHz
200 MHz to < 1 GHz
1 GHz to < 2 GHz
2 GHz to 6 GHz (TTR506A only above 3 GHz)
Uncorrected crosstalk with load -85 dB -110 dB -105 dB -120 dB -115 dB -120 dB Corrected crosstalk with load 1 – 90 dB -118 dB -115 dB -125 dB -125 dB -120 dB
1After full 2-port SOLT calibration with isolation using a Spinner BN533861 Type-N 50 Ω load.
Dynamic accuracy and compression
-
- Dynamic accuracy
-
Received level range 105 MHz 2 GHz > +5 to +10 dBm 0.65 dB 0.4 dB > 0 to +5 dBm 0.35 dB 0.2 dB > -20 to 0 dBm 0.2 dB 0.15 dB > -34 to -20 dBm 0.25 dB 0.2 dB > -50 to -34 dBm 0.35 dB 0.3 dB > -60 to -50 dBm 0.55 dB 0.45 dB - Dynamic accuracy, typical-mean
-
Received level range 10 MHz 105 MHz 350 MHz 783.5 MHz
1.083 GHz
2 GHz 3 GHz 4 GHz 5.25 GHz 6 GHz > +5 to +10 dBm 0.4 dB 0.3 dB 0.3 dB 0.3 dB 0.25 dB 0.2 dB 0.2 dB 0.15 dB 0.15 dB 0.15 dB > 0 to +5 dBm 0.25 dB 0.25 dB 0.1 dB 0.1 dB 0.1 dB 0.05 dB 0.05 dB 0.05 dB 0.05 dB 0.05 dB > -30 to 0 dBm 0.15 dB 0.1 dB 0.1 dB 0.1 dB 0.1 dB 0.1 dB 0.1 dB 0.1 dB 0.1 dB 0.1 dB -50 to -30 dBm 0.2 dB 0.15 dB 0.15 dB 0.15 dB 0.15 dB 0.15 dB 0.15 dB 0.15 dB 0.15 dB 0.15 dB
- Test port compression at maximum input level
- Compression at +10 dBm input level (typical-mean) is > +5 to +10 dBm
Frequency 10 MHz 105 MHz 350 MHz 783.5 MHz
1.083 MHz
2 GHz 3 GHz 4 GHz 5.25 GHz 6 GHz Compression 0.4 dB 0.4 dB 0.3 dB 0.25 dB 0.25 dB 0.2 dB 0.2 dB 0.2 dB 0.2 dB 0.2 dB
- Trace noise 1, typical
-
300 KHz to < 200 MHz 200 MHz to 6 GHz Magnitude 0.008 dB RMS 0.006 dB RMS Phase 0.05 degrees RMS 0.04 degrees RMS
1Measured in a 1 KHz IF BW, 10 dBm output power, thru connection
- Temperature stability 1, typical
-
300 KHz to 3 GHz > 3 GHz to 6 GHz Magnitude 0.008 dB/°C 0.015 dB/°C Phase 0.1 degrees/°C 0.2 degrees/°C
1Measured in a 10 Hz IF BW, 0 dBm output power, thru connection
- Maximum receiver input level
-
-
DC < 10 MHz ≥ 10 MHz to 6 GHz Operational ± 24 V 10 dBm 10 dBm Without damage ± 30 V 15 dBm 20 dBm
-
Output level calibration
-
Manufacturer Supported power sensors Tektronix USB power sensors PSM3000 series
PSM4000 series
PSM5000 seriesKeysight USB power sensors U848x series
U2020 series
U2000 seriesRohde and Schwarz USB power sensors NRP-Z power sensors
NRP-xxS/SN power sensors
Connectors
- Front panel
-
-
Name Function Description Port 1 VNA measurement port 1 Type N female conductor Port 2 VNA measurement port 2 Type N female conductor LO A LO A In/Out SMA female conductor LO B LO B In/Out SMA female conductor Power Power indicator Red/green LED
-
- Rear panel
-
-
Name Function Specifications Connector Bias 1 Port 1 bias tee connection ± 24 VDC
± 200 mA maximumBNC Female Bias 2 Port 2 bias tee connection Ref Out 10 MHz reference output 10 MHz ± 60 Hz
> 5 dBmRef In 10 MHz reference input 10 MHz ± 50 Hz
-5 dBm to + 10 dBmTrigger Trigger input TTL
± Edge, ± Level
> 50 nS durationAux Sync Auxiliary trigger TTL DC input Power supply 4.752 to 5.25 VDC 2.5 x 5.5 x 9.5 mm barrel connector
Center positiveUSB 2.0 USB connector for computer connection USB 2.0 USB 2.0 B connector
-
Power supply
-
Tektronix part number Input Output 119-8763-01 100 to 240 V AC, 50/60 Hz
IEC60320 C14 receptacle+ 5 V ±1%
4 A maximum
VectorVu-PC™ software
- System requirements
-
Processor Memory Drive Best Intel® Core™ i7 8 GB SSD Recommended Intel® Core™ i3 8 GB SSD Minimum Intel® Core™ i3 8 GB HDD Operating system: Windows® 7 or greater, 64-bit
Physical characteristics
- Length: 11.25” (28.58 cm)
Width: 8.125” (20.64 cm)
Depth: 1.75” (4.45 cm)
Weight: 3.5 lbs. (1.59 kg)
Environmental and safety
-
- Temperature
- Operating: +5 °C to +50 °C ( 41 °F to 122 °F)
Non-operating: -40 °C to 71 °C ( -40 °F to 159.8 °F)
- Humidity (operating)
- 5% to 80% ±5%RH (relative humidity) in the temperature range of +10 °C to 30 °C (+50 °F to 86 °F)
5% to 75% ±5%RH above +30 °C to 40 °C (+86 °F to 104 °F)
5% to 45% ±5%RH above +40 °C up to 50 °C (+104 °F to +122 °F)
Non-condensing
- Altitude
- Operating: 5000 m (16,404 ft)
Non-operating: 15,240 m (50,000 ft)
- Dynamics
-
- Vibration
- Operating: 0.31 GRMS: 5-500 Hz, 3 Axes at 10 min/axis
Non-operating: 2.46 GRMS: 5-500 Hz, 3 Axes at 10 min/axis
- Shock
- Operating: Half-sine mechanical shocks, 30 g peak amplitude, 11 msec duration, 3 drops in each direction of each axis, 18 total
Non-operating: Half-sine mechanical shocks, 40 g peak amplitude, 11 msec duration, 3 drops in each direction of each axis, 18 total
- Handling and transit
- Bench handling, operating: MIL-PRF-28800F Class 3
Transit drop, non-operating: MIL-PRF-28800F Class 2